Abstract Views :145 |
PDF Views:0
Authors
Affiliations
1 Matrix Laboratories Ltd., R and D Centre, Jinnaram Mandal, Medak-502325, IN
2 National Centre for Mass Spectrometry, Indian Institute of Chemical Technology, Hyderabad-500007, IN
3 Department of Chemistry, P.G. College of Science, Saifabad, Hyderabad, IN
Source
Asian Journal of Research in Chemistry, Vol 3, No 2 (2010), Pagination: 407-410
Abstract
A Liquid chromatographic (LC) method using mass spectrometric (MS) detection was developed and validated for the trace analysis (ppm level) of 4-Bromo methyl-2'-Cyanobiphenyl in Valsartan drug substances. LC analysis of 4-Bromo methyl-2'-Cyanobiphenyl was done on Necleosil C18 (100 mm×4.6 mm, 3 μm) column and the mobile phase is in the ratio of 50:50:1 containing water, acetonitrile, and acetic acid. The flow rate was 1.0 mL min-1 and the elution was monitored at 225 nm. The method was validated as per International Conference on Harmonization (ICH) guidelines. LC-MS is able to quantitate up to 1.0 ppm of 4-Bromo methyl-2'-cyanobiphenyl.
Keywords
Liquid Chromatography, Mass Spectrometry Valsartan.